Invacare Mobility Aid 1095340 User Manual

Solara Contracture Footplate Retro Fit Kit  
Assembly,Installation and Operating Instructions  
Kit Nos. 1095340, 1095341, 1095342, 1095343, 1095344, 1095345, 1095346  
SAVE THESE INSTRUCTIONS  
3. Remove the back canes. Refer to REMOVING/IN-  
NOTE: Check all parts for shipping damage. In case of  
damage, DO NOT use. Contact Carrier/Invacare for fur-  
ther instructions.  
STALLING THE BACK CANES in PROCEDURE 5  
of the Service Manual, part number 1085787.  
4. Remove the allen screw, spacer and locknut that  
secure the link bracket to the lower seat frame as  
shown in DETAIL “A” of FIGURE 1.  
SAFETY SUMMARY  
The following recommendations are made for the safe  
and proper installation and use of the Contracture  
Footplate Retro Fit Kit:  
5. Remove the shoulder screw and locknut securing  
the rear of the seat frame to the bearing housing as  
shown in DETAIL “A” of FIGURE 1.  
GENERAL WARNINGS  
6. Repeat STEPS 4-5 for opposite side of seat frame.  
7. Lift up to remove the seat frame from the base frame.  
DO NOT insta ll this e quipme nt without first re a d-  
ing a nd understa nding this instruction sheet. If you  
a re una ble to unde rsta nd the Wa rnings, Ca utions  
a nd Instruc tions, c onta c t a he a lthc a re profe s-  
siona l, dea ler or technica l personnel if a pplica ble,  
be fore a tte mpting to insta ll this e quipme nt - oth-  
e rwise , injury or da ma ge ma y oc c ur.  
NOTE: When reassembling the seat frame, the front  
crossbrace will be replaced with the crossmember as-  
sembly (insert tube and outer tube) provided in this kit  
andwillbesecuredwiththeEXISTINGmountingscrews.  
NOTE: Right and left sides of the seat frame are deter-  
mined by sitting in the wheelchair.  
INSTALLATION WARNINGS  
Afte r ANY a djustme nts, re pa ir or se rvic e a nd BE-  
FORE use , ma ke sure tha t a ll a tta c hing ha rdwa re  
is tighte ne d se c ure ly.  
8. On the left hand side of the seat frame, remove the  
two(2)EXISTINGmountingscrewsandlocknutsthat  
secure the front and rear crossbraces to the brace  
supports of the seat frame as shown in DETAIL “B”  
of FIGURE 1.  
This kit contains the following:  
DESCRIPTION  
QTY  
9. Pull the left hand side of the seat frame away from  
the front and rear crossmembers.  
Contracture Assembly (Right Side)  
Contracture Assembly (Left Side)  
Insert Tube (Crossmember Assembly)  
Outer Tube (Crossmember Assembly)  
1
1
1
1
10. Remove the EXISTING mounting screw, and locknut  
that secures the other end of the front crossbrace to  
the front right hand side brace support of the seat  
frame as shown in DETAIL “B” of FIGURE 1.  
INSTALLING THE CONTRACTURE  
FOOTPLATE ASSEMBLY  
Installing the Crossm em ber Assem bly  
(FIGURE 1)  
11. Remove the front crossbrace from the brace support  
and discard.  
12. Insert one end of the insert tube into the front right  
hand side brace support.  
NOTE: Refer to the GENERAL WARNINGS and INSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
13. Align mounting holes in the end of the insert tube with  
the brace support and secure with the EXISTING  
mounting screw and locknut.  
NOTE: Invacare recommends that two (2) people per-  
form this procedure.  
14. Slide the outer tube over the insert tube as shown in  
DETAIL “B” of FIGURE 1.  
NOTE:Takenoteofpositionandorientationofthemount- 15. Align the brace supports of the left hand side of the  
ing hardware for reassembly.  
seat frame with the front and rear crossmembers.  
16. Insert the front and rear crossmembers into the brace  
supportsofthelefthandsideoftheseatframe. Reinstall  
the two (2) EXISTING mounting screws and locknuts.  
1. If necessary, remove the seating system from the  
wheelchair. Refer to the seating systems Owner’s  
Manual supplied with the wheelchair.  
2. Remove the seat pan. Refer to REMOVING/IN-  
STALLING THE SEAT PAN in PROCEDURE 4 of  
the Service Manual, part number 1085787.  
1
 
CONTRACTURE ASSEMBLY ANGLE  
ADJ USTMENT (FIGURE 4)  
Angle Adjustment  
Bracket  
NOTE:RefertotheGENERALWARNINGSandINSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
Pop Rivet  
Button  
Quick Release Pin  
1. Remove the mounting screw and locknut used to lock  
the angle adjustment of the contracture assembly.  
Contracture  
Support  
2. Lift the contracture support UP/DOWN to achieve  
the desired angle.  
FIGURE 5 - USING THE OPTIONAL QUICK  
RELEASE PIN  
3. Align the mounting hole in the contracture support  
with one (1) of seven (7) mounting holes on the angle  
adjustment bracket.  
4. Reinstallthemountingscrewandlocknutusedtolock  
the angle adjustment of the contracture assembly.  
Securely tighten.  
ANGLE ADJ USTMENT OF THE  
CONTRACTURE FOOTPLATE  
(FIGURE 6)  
5. If necessary repeat STEPS 1-4 on remaining con-  
tracture assembly.  
NOTE: Refer to the GENERAL WARNINGS and INSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
Mounting Screw  
Angle  
Adjustment  
Bracket  
1. Loosen, but DO NOT, remove the two (2) flat screws,  
washers and locknuts that secure contracture  
footplate to the footplate hinge.  
Mounting  
Hole  
2. Tilt the contracture footplate UP/DOWN to achieve  
desired angle.  
Locknut  
3. Retighten the two (2) flat screws, washers and lock-  
nuts that secures the contracture foot plate to the  
footplate hinge.  
Contracture Support  
FIGURE 4- CONTRACTURE ASSEMBLY ANGLE  
ADJUSTMENT  
4. If necessary repeat STEPS 1-3 on remaining con-  
tracture footplate.  
USING THE OPTIONAL QUICK  
RELEASE PIN (FIGURE 5)  
Flat Screws  
Contracture  
Assembly  
NOTE: Refer to the GENERAL WARNINGS and INSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
Contracture  
Footplate  
NOTE: The Quick Release Pin replaces the mounting  
screw that is used to lock the angle adjustment of the  
contracture support.  
Footplate  
Hinge  
NOTE:Ifnotinstalledat time of shipment, thequickrelease  
pin must be installed by a qualified technician only, using a  
1/8-inch pop rivet.  
NOTE:Mountinghardware  
shown removed for clarity  
only.  
Washers  
Lock Nuts  
1. Press the button on the quick release pin and insert  
through the contracture support and the angle ad-  
justment bracket.  
FIGURE 6 - ANGLE ADJUSTMENT OF THE  
CONTRACTURE FOOTPLATE  
2. If necessary adjust the angle of the contracture as-  
sembly. Refer to CONTRACTURE ASSEMBLY  
ANGLE ADJUSTMENT in this instruction sheet.  
3. Repeat STEPS 1-2 on the remaining contracture as-  
sembly.  
3
 
4. Secure the footplate bracket to the contracture sup-  
port with the mounting screw, two (2) coved washers,  
washer and locknut. Securely tighten.  
CONTRACTURE FOOTPLATE  
DEPTH ADJ USTMENT (FIGURE 7)  
NOTE: Refer to the GENERAL WARNINGS and INSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
5. If necessary repeat STEPS 1-4 on remaining con-  
tracture footplate.  
Contracture Support  
Mounting  
Holes  
NOTE: Observe the angle of the contracture footplate for  
reinstallation.  
1. Remove the two (2) flat screws, washers and lock-  
nuts that secure the contracture footplate to the  
footplate hinge.  
Mounting Screws  
2. Move the contracture footplate to one (1) of four (4)  
mounting hole.  
Lock Nut  
Coved Washers  
Washers  
3. Reinstall the two (2) flat screws, washers and lock-  
nuts that secures the contracture foot plate to the  
footplate hinge.  
Footplate  
Bracket  
Flat Screws  
FIGURE 8 - CONTRACTURE FOOTPLATE HEIGHT  
ADJUSTMENT  
Mounting  
Hole  
REPLACING THE CONTRACTURE  
FOOTPLATE HEEL LOOP (FIGURE 9)  
Contracture  
Footplate  
NOTE:RefertotheGENERALWARNINGSandINSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
Washers  
Footplate Hinge  
1. Tilt the contracture footplate forward to gain access to the  
mounting screws that secure the heel loop to the under-  
side of the contracture foootplate. Refer to angle adjust-  
ment of the contracture footplate in this instruction sheet.  
Lock Nuts  
FIGURE 7 - CONTRACTURE FOOTPLATE DEPTH  
ADJUSTMENT  
2. Remove the four (4) mounting screws and washers  
that secure the EXISTING heel loop to the contrac-  
ture footplate.  
CONTRACTURE FOOTPLATE  
HEIGHT ADJ USTMENT (FIGURE 8)  
3. Reverse STEPS 1-2 to install the NEW heel loop.  
Contracture  
Footplate  
Washers  
NOTE: Refer to the GENERAL WARNINGS and INSTAL-  
LATION WARNINGS in the SAFETY SUMMARY of this  
instruction sheet.  
Mounting  
Screws  
1. Remove the mounting screw, two (2) coved washer,  
washer and locknut that secure the footplate bracket  
to the contracture support.  
Washers  
Heel Loop  
2. Move the footplate bracket UP/DOWN to achieve  
desired height.  
3. Align the mounting hole in the footplate bracket with  
one (1) of seven (7) mounting holes on the contrac-  
ture support.  
FIGURE 9 - REPLACING THE CONTRACTURE  
FOOTPLATE HEEL LOOP  
Invacare Corporation  
USA  
Canada  
One InvacareWay  
Elyria,Ohio USA  
44036-2125  
5970 ChedworthWay  
Mississauga, Ontario  
L5R 3T9, Canada  
905-890-8838  
Invacare and "Yes,you can" are trademarks of Invacare  
Corporation.  
© 2000 Invacare Corporation  
800-333-6900  
Form No.00-97  
Part No. 1095349 RevA (1) 5/00  
 

Indesit Refrigerator IN TSZ 1610 UK User Manual
Indesit Washer PWC 8148 W User Manual
JBL Car Speaker 552 User Manual
JBM electronic Network Router C20 through C21 User Manual
Jenn Air Cooktop CCE3401 User Manual
JVC Camcorder Accessories ABDIONICPKG User Manual
JVC CD Player XM EX90 User Manual
JVC Security Camera TK T2101RE User Manual
JVC Stereo System CA MXS6MD User Manual
Kenwood Speaker KFC P722 User Manual